型号: PCIe-6346

产品编号: 785813-01

SPECIFICATIONS PCIe-6346

500 kS/s/ch, 8 simultaneous AI, 24 DIO, 2 AO Multifunction I/O Device

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

         Typical specifications describe the performance met by a majority of models.

        





Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.


Specifications are Typical unless otherwise noted. Conditions

Specifications are valid for 25 °C unless otherwise noted.

Analog Input

Note Floating inputs can cause unnecessary power consumption and higher operating temperatures. NI recommends connecting unused analog input channels to

AIGND.

Number of channels                                           8 differential

ADC   resolution

16 bits

DNL

No missing codes, warranted

INL

Refer to the AI Absolute   Accuracy section.

Sample rate (simultaneous sampling on all channels sampled)

Maximum

500 kS/s

Minimum

No minimum

Timing   resolution

10 ns

Timing accuracy                                                50 ppm of sample rate

Input   coupling

DC

Input   range

±1 V, ±2 V, ±5 V, ±10 V

Maximum working voltage for all analog inputs (AI±)

Ranges   ±10 V, ±5 V

±11 V, Measurement Category I

Ranges   ±2 V, ±1 V

±9 V, Measurement Category I

         Caution Do not use for measurements within Categories II, III, and IV.

CMRR (at 60 Hz)                                               80 dB

Bandwidth   (small signal)

2.0 MHz at   ±1 V

2.9 MHz at other ranges

Input   impedance


Device   on


AI+ to   AI GND

>1 GΩ in parallel with 18 pF

AI- to   AI GND

>1 GΩ in parallel with 18 pF

Device   off


AI+ to   AI GND

2.37 kΩ

AI- to   AI GND

2.37 kΩ

Input   bias current

±6 nA

±90 nA, maximum over full   temperature range

Crosstalk   (at 100 kHz)

-80 dB

Input   FIFO size

4,095 samples shared among channels used

Data   transfers

DMA (scatter-gather), programmed I/O

Overvoltage   protection for AI<0..7>


Device   on

±30 V

Device   off

±15 V

Input   current during overvoltage conditions

±6.3 mA maximum/AI pin

Maximum   AI channels in overvoltage

4

Notice Exceeding overvoltage specifications may result in data corruption on nonovervoltaged channels.

Analog Triggers

Number of triggers                                           1

Source

AI <0..7>, APFI 0

Functions

Start   Trigger, Reference Trigger, Pause

Trigger,   Sample Clock, Sample Clock

Timebase

Source   level


AI   <0..7>

±Full scale

APFI 0

±10 V

Resolution

16 bits

Modes

Analog edge triggering, analog edge triggering with hysteresis,   and analog window triggering

Bandwidth   (large signal, to -3 dB)


AI   <0..7>

600 kHz

APFI 0

3.9 MHz

Accuracy

±1% of range

APFI 0   characteristics


Input   impedance

10 kΩ

Coupling

DC

Protection,   power on

±30 V

Protection,   power off

±15 V

AI Absolute Accuracy

Table 1. AI Absolute Accuracy

Nominal

Range

Positive Full   Scale

Nominal

Range

Negative Full

Scale

Residual

Gain Error

(ppm of

Reading)

Offset

Tempco

(ppm of

Range/°C)

Random

Noise, σ

Vrms)

Absolute

Accuracy   at

Full Scale

(μV)

10

-10

115

2

265

3,225

5

-5

115

2

148

1,613

Table 1. AI Absolute Accuracy (Continued)

Nominal

Range

Positive Full   Scale

Nominal

Range

Negative Full

Scale

Residual

Gain Error

(ppm of

Reading)

Offset

Tempco

(ppm of

Range/°C)

Random

Noise, σ

Vrms)

Absolute

Accuracy   at

Full Scale

(μV)

2

-2

117

2

74

650

1

-1

124

3

50

333

Note For more information about absolute accuracy at full scale, refer to the AI Absolute Accuracy Example section.

Gain tempco                                                      16.7 ppm/°C

Reference   tempco

5 ppm/°C

Residual   offset error

12 ppm of range

INL error

126 ppm of range

Note Accuracies listed are warranted for up to one year from the device external calibration when the device is within 10 °C of the external calibration temperature and 1 °C of the last self calibration, when averaging 10,000 DC samples. Other accuracies may be calculated for different temperatures and sample sizes using the given equations.

AI Absolute Accuracy Equation

AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainty

GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) +

ReferenceTempco · (TempChangeFromLastExternalCal)

OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INLError

                                         Random Noise ⋅   3

NoiseUncertainty =  for a coverage factor of 3 σ and averaging 10, 000 10,000 points.

AI Absolute Accuracy Example

Absolute accuracy at full scale on the analog input channels is determined using the following assumptions:

         TempChangeFromLastExternalCal = 10 °C

         TempChangeFromLastInternalCal = 1 °C

         number_of_readings = 10,000

         CoverageFactor = 3 σ

For example, on the 10 V range, the absolute accuracy at full scale is as follows:

GainError = 115 ppm + 16.7 ppm · 1 + 5 ppm · 10 = 181.7 ppm OffsetError = 12 ppm + 2 ppm ·1 + 126 ppm = 140 ppm

Noise Uncertainty =  = 8 µV

AbsoluteAccuracy = 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainty = 3225 µV

Analog Output

Number   of channels

2

DAC   resolution

16 bits

DNL

±1 LSB, maximum

Monotonicity

16 bit guaranteed

Accuracy

Refer to the AO Absolute   Accuracy section.

Maximum   update rate (simultaneous)


1   channel

900 kS/s

2   channels

840 kS/s

Minimum   update rate

No minimum

Timing   accuracy

50 ppm of sample rate

Timing   resolution

10 ns

Output   range

±10 V

Output   coupling

DC

Output   impedance

0.2 Ω

Output   current drive

±5 mA

Overdrive   protection

±15 V

Overdrive   current

15 mA

Power-on   state

±20 mV

Power-on/off   glitch

2 V peak for 500 ms

Output   FIFO size

8,191 samples shared among channels used

Data   transfers

DMA (scatter-gather), programmed I/O

AO   waveform modes

Non-periodic waveform, periodic waveform regeneration mode from   onboard FIFO, periodic waveform regeneration from host buffer including   dynamic update

Settling time, full-scale step,                               6 µs

15 ppm (1 LSB)

Slew   rate

15 V/µs

Glitch   energy at midscale transition

100 mV · 2.6 µs

AO Absolute Accuracy

Accuracies listed are warranted for up to one year from the device external calibration when the device is within 10 °C of the external calibration temperature and 1 °C of the last self calibration.

Table 2. AO Absolute Accuracy

Nominal

Range

Positive

Full

Scale

Nominal Range

Negative

Full

Scale

Residual

Gain

Error

(ppm   of

Reading)

Gain

Tempco

(ppm/°C)

Reference

Tempco

(ppm/°C)

Residual

Offset

Error

(ppm   of

Range)

Offset

Tempco

(ppm of Range/

°C)

INL

Error

(ppm of

Range)

Absolute

Accuracy at   Full

Scale

(μV)

10

-10

130

11.3

5

52

4.8

128

3,761

AO Absolute Accuracy Equation

AbsoluteAccuracy = OutputValue (GainError) + Range (OffsetError)

GainError = ResidualGainError + GainTempco (TempChangeFromLastInternalCal) + ReferenceTempco (TempChangeFromLastExternalCal)

OffsetError = ResidualOffsetError + OffsetTempco (TempChangeFromLastInternalCal) + INLError

Digital I/O/PFI

Static   Characteristics

Number   of channels

24 total

8   (P0.<0..7>)

16 (PFI <0..7>/P1, PFI <8..15>/P2)

Ground   reference

D GND

Direction   control

Each terminal individually programmable as input or output

Pull-down   resistor

50 kΩ,   typical 20 kΩ, minimum

Input   voltage protection

±20 V on up to two pins

      Caution Stresses beyond those listed under the Input voltage protection specification may cause permanent damage to the device.

Waveform Characteristics (Port 0 Only)

Terminals used                                                 Port 0 (P0.<0..7>)

Port/sample   size

Up to 8 bits

Waveform   generation (DO) FIFO

2,047 samples

Waveform   acquisition (DI) FIFO

255 samples

DI   Sample Clock frequency

0 to 1 MHz, system and bus   activity dependent

DO   Sample Clock frequency


Regenerate   from FIFO

0 to 1 MHz

Streaming   from memory

0 to 1 MHz, system and bus   activity dependent

Data   transfers

DMA (scatter-gather), programmed I/O

Digital   line filter settings

160 ns, 10.24 μs, 5.12 ms, disable

PFI/Port 1/Port 2 Functionality

Functionality                                                   Static digital input, static digital output, timing input, timing output

Timing   output sources

Many AI, AO, counter, DI, DO timing signals

Debounce   filter settings

90 ns, 5.12 µs, 2.56 ms, custom interval, disable; programmable   high and low transitions; selectable per input

Recommended Operating Conditions

Input   high voltage (VIH)


Minimum

2.2 V

Maximum

5.25 V

Input   low voltage (VIL)


Minimum

0 V

Maximum

0.8 V

Output   high current (IOH)


P0.<0..7>

-24 mA, maximum

PFI   <0..15>/P1/P2

-16 mA, maximum

Output low current (IOL)

P0.<0..7>

24 mA, maximum

PFI   <0..15>/P1/P2

16 mA, maximum

Digital I/O   Characteristics

Positive-going   threshold (VT+)

2.2 V, maximum

Negative-going   threshold (VT-)

0.8 V, minimum

Delta VT   hysteresis (VT+ - VT-)

0.2 V , minimum

IIL   input low current (VIN = 0 V)

-10 μA, maximum

IIH   input high current (VIN = 5 V)

250 μA, maximum

Figure 3. PFI <0..15>/P1/P2: IOH versus VOH

Figure 4. PFI <0..15>/P1/P2: IOL versus VOL

General-Purpose Counters

Number   of counter/timers

4

Resolution

32 bits

Counter   measurements

Edge counting, pulse, pulse width, semi-period, period, two-edge   separation

Position   measurements

X1, X2, X4   quadrature encoding with

Channel Z reloading; two-pulse encoding

Output   applications

Pulse, pulse train with dynamic updates, frequency division,   equivalent time sampling

Internal   base clocks

100 MHz, 20 MHz, 100 kHz

External   base clock frequency

0 MHz to 25 MHz

Base   clock accuracy

50 ppm

Inputs

Gate,   Source, HW_Arm, Aux, A, B, Z,

Up_Down, Sample Clock

Routing   options for inputs

Any PFI, RTSI, analog trigger, many internal triggers

FIFO

127 samples per counter

Data   transfers

Dedicated scatter-gather DMA controller for each counter/timer,   programmed I/O

Frequency Generator

Number   of channels

1

Base   clocks

20 MHz, 10 MHz, 100 kHz

Divisors

1 to 16

Base   clock accuracy

50 ppm

Output can be available on any PFI or RTSI terminal.

Phase-Locked Loop (PLL)

Table 3. Reference Clock Locking Frequencies

Reference Signal

PCI Express Locking Input   Frequency (MHz)

RTSI <0..7>

10, 20

PFI <0..15>

10, 20

Output of PLL                                                 100 MHz Timebase; other signals derived from

100 MHz Timebase including 20 MHz and

100 kHz Timebases

External Digital Triggers

Source

Any PFI, RTSI

Polarity

Software-selectable for most signals

Analog   input function

Start   Trigger, Reference Trigger, Pause

Trigger, Sample Clock, Convert Clock,   Sample

Clock Timebase

Analog   output function

Start   Trigger, Pause Trigger, Sample Clock,

Sample Clock Timebase

Counter/timer   functions

Gate,   Source, HW_Arm, Aux, A, B, Z,

Up_Down, Sample Clock

Digital waveform generation (DO) function

Start   Trigger, Pause Trigger, Sample Clock,

Sample Clock Timebase

Digital waveform acquisition (DI) function

Start Trigger, Reference Trigger, Pause

Trigger, Sample Clock, Sample Clock

Timebase

Device-to-Device Trigger Bus

Output selections                                               10 MHz Clock, frequency generator output;

many internal signals

Debounce   filter settings

90 ns, 5.12 μs, 2.56 ms, custom interval, disable; programmable   high and low transitions; selectable per input

Bus Interface


Form   factor

x4 PCI Express, specification v1.1 compliant

Slot   compatibility[1]

x4, x8, and x16 PCI Express or PCI Express hybrid slots

DMA   channels

8 (can be used for analog input, analog output, digital input,   digital output, counter/timer 0, counter/timer 1, counter/timer 2,   counter/timer 3)

Power Requirements

Caution The protection provided by the PCIe-6346 can be impaired if it is used in a manner not described in the user documentation.

Attention La protection apportée par le PCIe-6346 risque d'être endommagée s'il est utilisé d'une autre façon que celle décrite dans la documentation utilisateur.

+3.3 V                                                             2.1 W

+12 V

17.1 W

Current Limits


P0/PFI/P1/P2 and +5 V terminals  1.2 A, maximum combined

Physical Characteristics

Printed   circuit board dimensions

16.8 cm × 11.1 cm (6.60 × 4.38 in.)

PCI   Express mechanical form factor

Standard height, half length, single slot

Integrated   air mover (fan)

No

Weight

100 g (3.6 oz)

I/O   connectors


PCIe   device connector

68-Pos   Right Angle Single Stack PCB-Mount

VHDCI (Receptacle)

Cable   connector

68-Pos   Offset IDC Cable Connector (Plug)

(SHC68-*)

Note For more information about the connectors used for DAQ devices, refer to the document, NI DAQ Device Custom Cables, Replacement Connectors, and Screws, by going to ni.com/info and entering the Info Code rdspmb.

Calibration

Recommended   warm-up time

15 minutes

Calibration   interval

1 year

Maximum Working Voltage

Maximum working voltage refers to the signal voltage plus the common-mode voltage.

Channel to earth                                                 11 V, Measurement Category I

Caution Do not connect the PCIe-6346 to signals or use for measurements within Measurement Categories II, III, or IV.

Attention Ne connectez pas le PCIe-6346 à des signaux et ne l'utilisez pas pour effectuer des mesures dans les catégories de mesure II, III ou IV.

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated lowvoltage sources, and electronics.

Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Environmental Characteristics

Temperature   and Humidity


Temperature


Operating   (local operational ambient environment[2])

0 °C to 50 °C

Storage

-40 °C to 70 °C

Humidity

Operating

10% to 90% RH, noncondensing

Storage

5% to 95% RH, noncondensing

System   slot airflow

0.4 m/s (80 LFM)

Pollution   Degree

2

Maximum altitude                                           2,000 m (800 mbar) (at 25 °C ambient temperature)

Shock   and Vibration


Random   vibration


Operating

5 Hz to 500 Hz, 0.3 g RMS

Non-operating

5 Hz to 500 Hz, 2.4 g RMS

Operating   shock

30 g, half-sine, 11 ms pulse


 

PCIe-6346 Device Pinout


Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S.

Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2018 National Instruments. All rights reserved.

377315A-01    July 2, 2018



[1] Some motherboard reserve the x16 slot for graphics use. For PCI Express guidelines, refer to ni.com/pciexpress.

[2] For PCI Express adapter cards with integrated air movers, NI defines the local operational ambient environment to be at the fan inlet. For cards without integrated air movers, NI defines the local operational ambient environment to be 25 mm (1 in.) upstream of the leading edge of the card. For more information about the local operational ambient environment definition for PCI Express adapter cards, visit ni.com/info and enter the Info Code pcielocalambient.