型号: PXIe-6355

产品编号: 783632-01

DEVICE SPECIFICATIONS NI 6355

X Series Data Acquisition: 80 AI, 1.25 MS/s, 1 MS/s,






16-Bit Resolution, ±10 V, 24 DIO, 2 AO


Français

Deutsch

日本語

한국어

简体中文



ni.com/manuals



The following specifications are typical at 25 °C, unless otherwise noted. For more information about the NI 6355, refer to the X Series User Manual available at ni.com/manuals.

Analog Input

Number   of channels

40 differential or 80 single-ended

ADC   resolution

16 bits

DNL

No missing codes guaranteed

INL

Refer to the AI Absolute   Accuracy section.

Sample   rate


Single   channel maximum

1.25 MS/s

Multichannel   maximum

1.00 MS/s

Minimum

No minimum

Timing   resolution

10 ns

Timing   accuracy

50 ppm of sample rate

Input   coupling

DC

Input   range

±0.1 V, ±0.2 V, ±0.5 V, ±1   V, ±2 V, ±5 V, ±10 V

Maximum   working voltage for analog inputs

(signal   + common mode)

±11 V of AI GND

CMRR   (DC to 60 Hz)

100 dB

Figure 1. CMRR

                               10                        100

        1 k                       10 k

Frequency (Hz)

100 k

Input   impedance



Device   on



AI+ to   AI GND

>10   GΩ in parallel with 100 pF

AI- to   AI GND

>10   GΩ in parallel with 100 pF

Device   off


AI+ to   AI GND

820 Ω

AI- to   AI GND

820 Ω

Input   bias current

±100   pA

Crosstalk   (at 100 kHz)


Adjacent   channels

-75 dB

Non-adjacent   channels

-88 dB

Small   signal bandwidth (-3 dB)

1.7   MHz

Input   FIFO size

4,095   samples

Scan   list memory

4,095   entries

Data   transfers

DMA   (scatter-gather), programmed I/O

Overvoltage protection for all analog input and sense channels

Settling Time for Multichannel Measurements

Range

±60 ppm of Step   (±4 LSB for Full-Scale Step)

±15 ppm of Step   (±1 LSB for Full-Scale Step)

± 10 V, ±5   V, ±2 V,

±1 V

1 μs

1.5 μs

±0.5 V

1.5 μs

2 μs

±0.2 V, ±0.1 V

2 μs

8 μs

Figure 2. Settling Time versus Time for Different Source Impedances

AI Absolute Accuracy

Table 1. AI Absolute Accuracy

Nominal

Range

Positive

Full   Scale

Nominal Range

Negative

Full   Scale

Residual

Gain   Error

(ppm of

Reading)

Residual

Offset

Error (ppm of   Range)

Offset

Tempco

(ppm   of

Range/°C)

Random

Noise,   σ

Vrms)

Absolute

Accuracy at Full Scale[1]

V)

10

-10

48

13

21

281

1520

5

-5

55

13

21

137

800

2

-2

55

13

24

56

320

Table 1. AI Absolute Accuracy (Continued)


Nominal

Range

Positive

Full   Scale

Nominal Range

Negative

Full   Scale

Residual

Gain   Error

(ppm of

Reading)

Residual

Offset

Error (ppm of   Range)

Offset

Tempco

(ppm   of

Range/°C)

Random

Noise,   σ

Vrms)

Absolute

Accuracy at Full Scale[2]

V)


1

-1

65

17

27

35

180


0.5

-0.5

68

17

34

26

95


0.2

-0.2

95

27

55

21

50


0.1

-0.1

108

45

90

16

32

Gain   tempco

13 ppm/°C


Reference   tempco

1 ppm/°C


INL   error

46 ppm of range












Note Accuracies listed are valid for up to two years from the device external calibration.

AI Absolute Accuracy Equation

AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainty

GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) +

ReferenceTempco · (TempChangeFromLastExternalCal)

OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INLError

Random Noise ⋅ 3

NoiseUncertainty =  for a coverage factor of 3 σ and averaging 10,000 10, 000

points.

AI Absolute Accuracy Example

Absolute accuracy at full scale on the analog input channels is determined using the following assumptions:

         TempChangeFromLastExternalCal = 10 °C

         TempChangeFromLastInternalCal = 1 °C

         number_of_readings = 10,000

         CoverageFactor = 3 σ

For example, on the 10 V range, the absolute accuracy at full scale is as follows: GainError = 48 ppm + 13 ppm · 1 + 1 ppm · 10 = 71 ppm

OffsetError = 13 ppm + 21 ppm · 1 + 46 ppm = 80 ppm

NoiseUncertainity =  = 8.4 µV

AbsoluteAccuracy = 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainity = 1, 520 µV

Analog Triggers

Number   of triggers

1

Source

AI <0..79>

Functions

Start   Trigger, Reference Trigger, Pause

Trigger,   Sample Clock, Convert Clock,

Sample Clock Timebase

Source   level


AI   <0..79>

±Full scale

APFI 0

±10 V

Resolution

16 bits

Modes

Analog edge triggering, analog edge triggering with hysteresis,   and analog window triggering

Bandwidth   (-3 dB)


AI   <0..79>

3.4 MHz

APFI 0

3.9 MHz

Accuracy

±1% of range

APFI 0   characteristics


Input   impedance

10 kΩ

Coupling

DC

Protection,   power on

±30 V

Protection,   power off

±15 V

Analog Output


Number   of channels

2

DAC   resolution

16 bits

DNL                                                                   ±1 LSB

Monotonicty

16-bit guaranteed

Maximum   update rate


1   channel

2.86 MS/s

2   channels

2.00 MS/s

Timing   accuracy

50 ppm of sample rate

Timing   resolution

10 ns

Output   range

±10 V, ±5 V, ±external reference on APFI 0

Output   coupling

DC

Output   impedance

0.2 Ω

Output   current drive

±5 mA

Overdrive   protection

±25 V

Overdrive   current

26 mA

Power-on   state

±5 mV

Power-on/off   glitch

1.5 V peak for 200 ms

Output   FIFO size

8,191 samples shared among channels used

Data   transfers

DMA (scatter-gather), programmed I/O

AO   waveform modes

Non-periodic waveform, periodic waveform regeneration mode from   onboard FIFO, periodic waveform regeneration from host buffer including   dynamic update

Settling   time, full-scale step, 15 ppm (1 LSB)

2 µs

Slew   rate

20 V/µs

Glitch   energy at midscale transition, ±10 V range

10 nV · s

External   Reference


APFI 0   characteristics


Input   impedance

10 kΩ

Coupling

DC

Protection,   device on

± 30 V

Protection,   device off

± 15 V

Range

± 11 V

Slew   rate

20 V/μs

Figure 3. AO External Reference Bandwidth

AO Absolute Accuracy

Absolute accuracy at full-scale numbers is valid immediately following self calibration and assumes the device is operating within 10 °C of the last external calibration.

Table 2. AO Absolute Accuracy

Nominal Range

Positive Full   Scale

Nominal   Range

Negative Full

Scale

Residual Gain

Error (ppm of   Reading)

Gain Tempco   (ppm/°C)

Absolute

Accuracy   at Full

Scale (μV)

10

-10

63

17

1890

5

-5

70

8

935

Reference   tempco (ppm/°C)

1

Residual   offset error (ppm of range)

33

Offset   tempco (ppm of range/°C)

2

INL   Error (ppm of range)

64







Note Accuracies listed are valid for up to two years from the device external calibration.

AO Absolute Accuracy Equation

AbsoluteAccuracy = OutputValue · (GainError) + Range · (OffsetError)

GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) +

ReferenceTempco · (TempChangeFromLastExternalCal)

OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INLError

Digital I/O/PFI

Static   Characteristics


Number   of channels

24 total,   8 (P0.<0..7>),

16 (PFI <0..7>/P1, PFI <8..15>/P2)

Ground   reference

D GND

Direction   control

Each terminal individually programmable as input or output

Pull-down   resistor

50 kΩ typical, 20 kΩ minimum

Input   voltage protection

±20 V on up to two pins

      Caution Stresses beyond those listed under the Input voltage protection specification may cause permanent damage to the device.

Input   FIFO size

255 samples

Output   FIFO size

2,047 samples

Resolution

32 bits

Waveform Characteristics (Port 0 Only)

Terminals   used

Port 0 (P0.<0..7>)

Port/sample   size

Up to 8 bits

Waveform   generation (DO) FIFO

2,047 samples

Waveform   acquisition (DI) FIFO

255 samples

DI   Sample Clock frequency

0 to 10 MHz, system and bus activity dependent

DO   Sample Clock frequency

0 to 10 MHz, system and bus activity dependent

Data   transfers

DMA (scatter-gather), programmed I/O

Digital   line filter settings

160 ns, 10.24 μs, 5.12 ms, disable

PFI/Port 1/Port 2 Functionality

Functionality

Static digital input, static digital output, timing input, timing   output

Timing   output sources

Many AI, AO, counter, DI, DO timing signals

Debounce   filter settings

90 ns, 5.12 μs, 2.56 ms, custom interval, disable; programmable   high and low transitions; selectable per input

Recommended Operation Conditions

Input   high voltage (VIH)


Minimum

2.2 V

Maximum

5.25 V

Input   low voltage (VIL)


Minimum

0 V

Maximum

0.8 V

Output   high current (IOH)


P0.<0..7>

-24 mA maximum

PFI   <0..15>/P1/P2

-16 mA maximum

Output   low current (IOL)


P0.<0..7>

24 mA maximum

PFI   <0..15>/P1/P2

16 mA maximum

Digital I/O   Characteristics


Positive-going   threshold (VT+)

2.2 V maximum

Negative-going   threshold (VT-)

0.8 V minimum

Delta   VT hysteresis (VT+ - VT-)

0.2 V minimum

IIL   input low current (VIN = 0 V)

-10 μA maximum

IIH   input high current (VIN = 5 V)

250 μA maximum

Figure 6. PFI <0..15>/P1/P2: IOH versus VOH

Figure 7. PFI <0..15>/P1/P2: IOL versus VOL

                              0                 0.2                0.4

General-Purpose Count

    0.6               0.8                1.0                1.2

VOL (V)

ers

Number   of counter/timers

4

Resolution

32 bits

Counter measurements                                    Edge counting, pulse, pulse width, semi-period, period, two-edge separation

Position   measurements

X1, X2, X4   quadrature encoding with

Channel Z reloading; two-pulse encoding

Output   applications

Pulse, pulse train with dynamic updates, frequency division,   equivalent time sampling

Internal   base clocks

100 MHz, 20 MHz, 100 kHz

External   base clock frequency

0 MHz to   25 MHz; 0 MHz to 100 MHz on

PXIe_DSTAR <A,B>

Base   clock accuracy

50 ppm

Inputs

Gate,   Source, HW_Arm, Aux, A, B, Z,

Up_Down, Sample Clock

Routing   options for inputs

Any PFI, PXIe_DSTAR<A,B>, PXI_TRIG, PXI_STAR, analog   trigger, many internal signals

FIFO

127 samples per counter

Data   transfers

Dedicated scatter-gather DMA controller for each counter/timer,   programmed I/O

Frequency   Generator


Number   of channels

1

Base   clocks

20 MHz, 10 MHz, 100 kHz

Divisors

1 to 16

Base   clock accuracy

50 ppm

Phase-Locked Loop (PLL)

Table 3. Reference Clock Locking Frequencies

Reference Signal

PXI Express Locking Input   Frequency (MHz)

PXIe_DSTAR<A,B>

10, 20, 100

PXI_STAR

10, 20

PXIe_CLK100

100

Table 3. Reference Clock Locking Frequencies (Continued)

Reference Signal

PXI Express Locking Input   Frequency (MHz)

PXI_TRIG <0..7>

10, 20

PFI <0..15>

10, 20

Output of PLL                                                  100 MHz Timebase; other signals derived

from 100 MHz Timebase including 20 MHz and 100 kHz Timebases.

External Digital Triggers

Source

Any PFI,   PXIe_DSTAR<A,B>, PXI_TRIG,

PXI_STAR

Polarity

Software-selectable for most signals

Analog   input function

Start   Trigger, Reference Trigger, Pause

Trigger,   Sample Clock, Convert Clock,

Sample Clock Timebase

Analog   output function

Start   Trigger, Pause Trigger, Sample Clock,

Sample Clock Timebase

Counter/timer   functions

Gate,   Source, HW_Arm, Aux, A, B, Z,

Up_Down, Sample Clock

Digital   waveform generation (DO) function

Start   Trigger, Pause Trigger, Sample Clock,

Sample Clock Timebase

Digital   waveform acquisition (DI) function

Start   Trigger, Reference Trigger, Pause

Trigger,   Sample Clock, Sample Clock

Timebase

Device-to-Device Trigger Bus

Input   source

PXI_TRIG   <0..7>, PXI_STAR,

PXIe_DSTAR<A,B>

Output   destination

PXI_TRIG <0..7>, PXIe_DSTARC

Output   selections

10 MHz Clock, frequency generator output, many internal signals

Debounce   filter settings

90 ns, 5.12 μs, 2.56 ms, custom interval, disable; programmable   high and low transitions; selectable per input

Bus Interface

Form   factor

x1 PXI Express peripheral module, specification rev. 1.0 compliant

Slot   compatibility

x1 and x4 PXI Express or PXI Express hybrid slots

DMA   channels

8: analog input, analog output, digital input, digital output,   counter/timer 0, counter/timer

1, counter/timer 2, counter/timer 3

Devices may be installed in PXI Express slots or PXI Express hybrid slots.

Power Requirements

Caution The protection provided by the device can be impaired if the device is used in a manner not described in the X Series User Manual.

+3.3 V

1.6 W

+12 V

19.8 W

Current Limits


+5 V   terminal (connector 0)

1 A max

P0/PFI/P1/P2   and +5 V terminals combined

1.5 A max

Physical Characteristics

Device   dimensions

Standard 3U PXI

Weight

198 g (7.0 oz)

I/O   connectors

2 68-pin VHDCI

         Caution If you need to clean the module, wipe it with a dry towel.

Calibration (AI and AO)

Recommended   warm-up time

15 minutes

Calibration   interval

2 years

Maximum Working Voltage

Maximum working voltage refers to the signal voltage plus the common-mode voltage.

Channel to earth                                               11 V, Measurement Category I

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated lowvoltage sources, and electronics.

Caution Do not use for measurements within Categories II, III, or IV.

Note Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Shock and Vibration

Operational   shock

30 g peak, half-sine, 11 ms pulse (Tested in accordance with   IEC-60068-2-27. Test profile developed in accordance with

MIL-PRF-28800F.)

Random   vibration


Operating

5 to 500 Hz, 0.3 grms

Nonoperating

5 to 500 Hz, 2.4 grms (Tested in accordance with   IEC-60068-2-64. Nonoperating test profile exceeds the requirements of

MIL-PRF-28800F, Class 3.)

Environmental

Operating   temperature

0 to 55 ºC

Storage   temperature

-40 to 70 ºC

Operating   humidity

10 to 90% RH, noncondensing

Storage   humidity

5 to 95% RH, noncondensing

Pollution   degree

2

Maximum   altitude

2,000 m

Indoor   use only.

Safety


This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:

         IEC 61010-1, EN 61010-1

         UL 61010-1, CSA 61010-1

Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:

         EN 61326-1 (IEC 61326-1): Class B emissions; Basic immunity

         EN 55011 (CISPR 11): Group 1, Class B emissions

         EN 55022 (CISPR 22): Class B emissions

         EN 55024 (CISPR 24): Immunity

         AS/NZS CISPR 11: Group 1, Class B emissions

         AS/NZS CISPR 22: Class B emissions

         FCC 47 CFR Part 15B: Class B emissions

         ICES-001: Class B emissions

Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.

Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

Note For EMC declarations and certifications, and additional information, refer to the Online Product Certification section.

This product meets the essential requirements of applicable European Directives, as follows:

         2006/95/EC; Low-Voltage Directive (safety)

         2014/30/EU; Electromagnetic Compatibility Directive (EMC)

Online Product Certification

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

Waste Electrical and Electronic Equipment (WEEE)

    EU Customers At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee. 电子信息产品污染控制管理办法(中国RoHS

中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录

ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

Device Pinout

Figure 8. NI PXIe-6355 Pinout

AI 0 (AI 0+) 68 34 AI 8 (AI 0–) AI 71 (AI 71+) 1 35 AI 79 (AI 71–) AI GND 67 33 AI 1 (AI 1+) AI 78 (AI 70–) 2 36 AI 70 (AI 70+)

AI 9 (AI 1–)         66   32     AI GND                                                                  AI 69 (AI 69+)       3    37     AI 77 (AI 69–)

AI 2 (AI 2+) 65 31 AI 10 (AI 2–) AI 68 (AI 68+) 4 38 AI 76 (AI 68–) AI GND 64 30 AI 3 (AI 3+) AI 75 (AI 67–) 5 39 AI 67 (AI 67+)

AI 11 (AI 3–)       63   29     AI GND                                                                  AI 66 (AI 66+)       6    40     AI 74 (AI 66–)

AI SENSE           62   28     AI 4 (AI 4+)                                                            AI 65 (AI 65+)       7    41     AI 73 (AI 65–)

AI 12 (AI 4–) 61       27       AI GND          AI 72 (AI 64–)          8         42       AI 64 (AI 64+) AI 5 (AI 5+)  60       26       AI 13 (AI 5–) AI GND   9         43       AI GND

AI GND                59   25     AI 6 (AI 6+)                                                            AI 55 (AI 55+)      10   44     AI 63 (AI 55–)

AI 14 (AI 6–)       58   24     AI GND                                                                  AI 54 (AI 54+)      11    45     AI 62 (AI 54–)

AI 7 (AI 7+)         57   23     AI 15 (AI 7–)                                                          AI 61 (AI 53–)      12   46     AI 53 (AI 53+)

AI GND                56   22     AO 0                                                                      AI 52 (AI 52+)      13   47     AI 60 (AI 52–)

AO GND 55 21 AO 1 AI 51 (AI 51+) 14 48 AI 59 (AI 51–) AO GND 54 20 APFI 0 AI 58 (AI 50–) 15 49 AI 50 (AI 50+)

D GND                53   19     P0.4                                                                        AI 49 (AI 49+)      16   50     AI 57 (AI 49–)

P0.0 52 18 D GND AI 48 (AI 48+) 17 51 AI 56 (AI 48–) P0.5 51 17 P0.1 AI 47 (AI 39–) 18 52 AI 39 (AI 39+)

D GND                50   16     P0.6                                                                        AI 38 (AI 38+)      19   53     AI 46 (AI 38–)

P0.2                     49   15     D GND                                                                   AI 37 (AI 37+)      20   54     AI 45 (AI 37–)

P0.7                     48   14      +5 V                                                                       AI 44 (AI 36–)      21   55     AI 36 (AI 36+)

P0.3 47 13 D GND AI GND 22 56 AI SENSE 2 PFI 11/P2.3 46 12 D GND AI 35 (AI 35+) 23 57 AI 43 (AI 35–)

PFI 10/P2.2 45 11 PFI 0/P1.0 AI 34 (AI 34+) 24 58 AI 42 (AI 34–) D GND 44 10 PFI 1/P1.1 AI 41 (AI 33–) 25 59 AI 33 (AI 33+)

PFI 2/P1.2           43    9      D GND                                                                   AI 32 (AI 32+)      26   60     AI 40 (AI 32–)

PFI 3/P1.3 42 8 +5 V AI 23 (AI 23+) 27 61 AI 31 (AI 23–) PFI 4/P1.4 41 7 D GND AI 30 (AI 22–) 28 62 AI 22 (AI 22+)

PFI 13/P2.5        40    6      PFI 5/P1.5                                                             AI 21 (AI 21+)      29   63     AI 29 (AI 21–)

PFI 15/P2.7 39 5 PFI 6/P1.6 AI 20 (AI 20+) 30 64 AI 28 (AI 20–) PFI 7/P1.7 38 4 D GND AI 27 (AI 19–) 31 65 AI 19 (AI 19+)

PFI 8/P2.0           37    3      PFI 9/P2.1                                                             AI 18 (AI 18+)      32   66     AI 26 (AI 18–)

D GND                36    2      PFI 12/P2.4                                                           AI 17 (AI 17+)      33   67     AI 25 (AI 17–)

D GND                35    1      PFI 14/P2.6                                                           AI 24 (AI 16–)      34   68     AI 16 (AI 16+)


Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.

U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2014—2015 National Instruments. All rights reserved.

374653C-01     Sep15



[1] Refer to the AI Absolute Accuracy Example section.

[2] Refer to the AI Absolute Accuracy Example section.