DEVICE SPECIFICATIONS NI 6230
M Series Data Acquisition: 16-Bit, 250 kS/s, 8 AI, 4 AO, 6 DI, 4 DO, Voltage Input/Output, Bank Isolation
The following specifications are typical at 25 °C, unless otherwise noted. For more information about the NI 6230, refer to the NI 6230 User Manual available at ni.com/manuals.
Number of channels | 4 differential or 8 single ended |
Channel type | Voltage input |
ADC resolution | 16 bits |
DNL | No missing codes guaranteed |
INL | Refer to the AI Absolute Accuracy section |
Sample rate | |
Single channel maximum | 250 kS/s |
Minimum | 0 S/s |
Timing resolution | 50 ns |
Timing accuracy | 50 ppm of sample rate |
Input coupling | DC |
Input range | ±0.2 V, ±1 V, ±5 V, ±10 V |
Maximum working voltage for analog inputs | Refer to the Maximum Working Voltage section |
CMRR (DC to 60 Hz) | 95 dB (with respect to AI GND) |
Input impedance | |
Device on | |
AI+ to AI GND | >10 GΩ in parallel with 100 pF |
AI- to AI GND | >10 GΩ in parallel with 100 pF |
Device off
AI+ to AI GND | 820 Ω |
AI- to AI GND | 820 Ω |
Input bias current | ±100 pA |
Crosstalk (at 100 kHz) | |
Adjacent channels | -75 dB |
Non-adjacent channels | -90 dB |
Small signal bandwidth (-3 dB) | 700 kHz |
Input FIFO size | 4,095 samples |
Scan list memory | 4,095 entries |
Data transfers | DMA (scatter-gather), interrupts, programmed I/O |
Overvoltage protection (AI <0..7> with respect to AI GND)
Accuracy, full scale step, all ranges | |
±90 ppm of step (±6 LSB) | 4 µs convert interval |
±30 ppm of step (±2 LSB) | 5 µs convert interval |
±15 ppm of step (±1 LSB) | 7 µs convert interval |
Figure 1. Settling Error versus Time for Different Source Impedances
Figure 2. AI Small Signal Bandwidth
Figure 3. AI CMRR to Earth Ground
Figure 4. AI CMRR to AI GND
Note Accuracies listed are valid for up to one year from the device external calibration.
Table 1. AI Absolute Accuracy
Nominal Range Positive Full Scale | Nominal Range Negative Full Scale | Residual Gain Error (ppm of Reading) | Residual Offset Error (ppm of Range) | Offset Tempco (ppm of Range/°C) | Random Noise, σ (μVrms) | Absolute Accuracy at Full Scale (μV) | Sensitivity (μV) |
10 | -10 | 75 | 20 | 57 | 244 | 3,100 | 97.6 |
5 | -5 | 85 | 20 | 60 | 122 | 1,620 | 48.8 |
1 | -1 | 95 | 25 | 79 | 30 | 360 | 12.0 |
0.2 | -0.2 | 135 | 80 | 175 | 13 | 112 | 5.2 |
Note Sensitivity is the smallest voltage change that can be detected. It is a function of noise.
Gain tempco 25 ppm/°C
Reference tempco | 5 ppm/°C |
INL error | 76 ppm of range |
AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainty
GainError = ResidualAIGainError + GainTempco · (TempChangeFromLastInternalCal)
+ ReferenceTempco · (TempChangeFromLastExternalCal)
OffsetError = ResidualAIOffsetError + OffsetTempco ·
(TempChangeFromLastInternalCal) + INLError
Random Noise ⋅ 3
NoiseUncertainty = for a coverage factor of 3 σ and averaging
100 points.
Absolute accuracy at full scale on the analog input channels is determined using the following assumptions:
• TempChangeFromLastExternalCal = 10 °C
• TempChangeFromLastInternalCal = 1 °C
• number_of_readings = 100
• CoverageFactor = 3 σ
For example, on the 10 V range, the absolute accuracy at full scale is as follows:
GainError = 75 ppm + 25 ppm · 1 + 5 ppm · 10 = 150 ppm
OffsetError = 20 ppm + 57 ppm · 1 + 76 ppm = 153 ppm
NoiseUncertainity = = 73 µV
AbsoluteAccuracy = 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainity = 3,100 µV
Number of channels | 4 |
Channel type | Voltage output |
DAC resolution | 16 bits |
DNL | ±1 LSB |
Monotonicity | 16 bit guaranteed |
Maximum update rate | |
1 channel | 500 kS/s |
2 channels | 450 kS/s per channel |
3 channels | 425 kS/s per channel |
4 channels | 400 kS/s per channel |
Timing accuracy | 50 ppm of sample rate |
Timing resolution | 50 ns |
Output range | ±10 V |
Output coupling | DC |
Output impedance | 0.4 Ω |
Output current drive | ±5 mA |
Overdrive protection | ±25 V |
Overdrive current | 10 mA |
Power-on state | ±20 mV |
Power-on glitch[1] | ±2 V for 2 ms |
Power-off glitch[2] | ±100 mV for 350 ms |
Output FIFO size | 8,191 samples shared among channels used |
Data transfers | DMA (scatter-gather), interrupts, programmed I/O |
AO waveform modes Non-periodic waveform, periodic waveform
regeneration mode from onboard FIFO, periodic waveform regeneration from host buffer including dynamic update
Settling time, full-scale step, 15 ppm (1 LSB) | 6 µs |
Slew rate | 15 V/µs |
Glitch energy | |
Magnitude | 100 mV |
Duration | 3 µs |
Absolute accuracy at full-scale numbers is valid immediately following internal calibration and assumes the device is operating within 10 °C of the last external calibration.
Note Accuracies listed are valid for up to one year from the device external calibration.
Table 2. AO Absolute Accuracy
Nominal Range Positive Full Scale (V) | Nominal Range Negative Full Scale (V) | Residual Gain Error (ppm of Reading) | Gain Tempco (ppm/°C) | Residual Offset Error (ppm of Range) | Offset Tempco (ppm of Range/°C) | Absolute Accuracy at Full Scale (μV) |
10 | -10 | 90 | 10 | 40 | 5 | 3,230 |
Reference tempco
INL error
AbsoluteAccuracy = OutputValue · (GainError) + Range · (OffsetError)
GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) +
ReferenceTempco · (TempChangeFromLastExternalCal)
OffsetError = ResidualOffsetError + AOOffsetTempco ·
(TempChangeFromLastInternalCal) + INLError
Static Characteristics Number of channels | 10 total |
Number of input channels | 6 (PFI <0..5>/P0.<0..5>) |
Number of output channels | 4 (PFI <6..9>/P1.<0..3>) |
Ground reference | D GND |
Direction control | Fixed, lines are unidirectional |
Input voltage protection | ±20 V on up to two pins[3] |
PFI <0..5>/P0.<0..5> Static digital input, timing input
PFI <6..9>/P1.<0..3> | Static digital output, timing output |
Timing output sources | Many AI, AO, counter timing signals |
Debounce filter settings | 125 ns, 6.425 µs, 2.56 ms, disable; high and low transitions; selectable per input |
Level | Min | Max |
VIL input low voltage | 0 V | 0.8 V |
VIH input high voltage | 2 V | 5.25 V |
IIL input low current (Vin = 0 V) | — | -10 µA |
IIH input high current (Vin = 5 V) | — | 10 µA |
Table 3. Guaranteed Output Levels
Level | Voltage Level | Current Level |
VOL | 0.4 V | 7 mA |
VOL | 0.6 V | 10 mA |
Table 3. Guaranteed Output Levels (Continued)
Level | Voltage Level | Current Level |
VOH | 2.8 V | -24 mA |
VOH | 4.0 V | -6 mA |
Level | Minimum | Maximum |
IOL output low current P1.<0..3> | — | 10 mA |
IOH output high current P1.<0..3> | — | -24 mA |
Number of counter/timers | 2 |
Resolution | 32 bits |
Counter measurements | Edge counting, pulse, semi-period, period, two-edge separation |
Position measurements | X1, X2, X4 quadrature encoding with Channel Z reloading; two-pulse encoding |
Output applications | Pulse, pulse train with dynamic updates, frequency division, equivalent time sampling |
Internal base clocks | 80 MHz, 20 MHz, 0.1 MHz |
External base clock frequency | 0 MHz to 20 MHz |
Base clock accuracy | 50 ppm |
Inputs | Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down |
Routing options for inputs | Any input PFI, RTSI, PXI_TRIG, PXI_STAR, many internal signals |
FIFO | 2 samples |
Data transfers | Dedicated scatter-gather DMA controller for each counter/timer; interrupts; programmed I/O |
Maximum propagation delay through isolator | |
Digital inputs | 35 ns |
Digital outputs | 45 ns |
Propagation delay skew between channels (inputs and outputs) | 15 ns |
Frequency Generator | |
Number of channels | 1 |
Base clocks | 10 MHz, 100 kHz |
Divisors | 1 to 16 |
Base clock accuracy | 50 ppm |
Output can be available on any output PFI or RTSI terminal.
Number of PLLs | 1 |
Reference signal | PXI_STAR, PXI_CLK10, RTSI <0..7> |
Output of PLL | 80 MHz Timebase; other signals derived from 80 MHz Timebase including 20 MHz and 100 kHz Timebases |
Source | Any PFI, RTSI, PXI_TRIG, PXI_STAR |
Polarity | Software-selectable for most signals |
Analog input function | Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert Clock, Sample Clock Timebase |
Analog output function | Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase |
Counter/timer function | Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down |
PCI RTSI <0..7>[4]
PXI | PXI_TRIG <0..7>, PXI_STAR |
Output selections | 10 MHz Reference Clock, frequency generator output, many internal signals |
Debounce filter settings | 125 ns, 6.425 μs, 2.56 ms, disable; high and low transitions; selectable per input |
Bus Interface | |
PCI or PXI | 3.3 V or 5 V signal environment |
PXI-6230 devices can be installed in PXI slots or PXI Express hybrid slots.
DMA channels 4, analog input, analog output, counter/timer 0, counter/timer 1
Current draw from bus during no-load condition
+5 V | 0.5 A |
+12 V | 20 mA |
Current draw from bus during AI and AO overvoltage condition
+5 V | 0.75 A |
+12 V | 20 mA |
Printed circuit board dimensions | |
PCI | 9.7 cm × 15.5 cm (3.8 in. × 6.1 in.) |
PXI | Standard 3U PXI |
Weight | |
PCI | 110 g (3.8 oz) |
PXI | 150 g (5.2 oz) |
I/O connector 37-pin D-SUB
Calibration | |
Recommended warm-up time | 15 minutes |
Calibration interval | 1 year |
Maximum working voltage refers to the signal voltage plus the common-mode voltage.
Channel-to-earth ground[5] | |
dielectric withstand test | |
Channel-to-bus6 | |
dielectric withstand test | |
Analog channel-to-AI GND or AO GND (in the following figure, |Va - Vc|) | ≤11 V, Measurement Category I |
Digital channel-to-D GND (in the following figure, |Vb - Vc|) | ≤5.25 V, Measurement Category I |
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated lowvoltage sources, and electronics.
Caution This device is rated for Measurement Category I and the voltage across the isolation barrier is limited to no greater than 30 Vrms/60 VDC/42.4 Vpk continuous. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.
The following figure illustrates the maximum working voltage specifications.
Figure 5. NI 6230 Maximum Working Voltage
Operating environment | |
Ambient temperature range | 0 to 55 °C (Tested in accordance with IEC-60068-2-1 and IEC-60068-2-2.) |
Relative humidity range | 10% to 90%, noncondensing (Tested in accordance with IEC-60068-2-56.) |
Storage environment | |
Ambient temperature range | -40 to 70 °C (Tested in accordance with IEC-60068-2-1 and IEC-60068-2-2.) |
Relative humidity range | 5% to 95% noncondensing (Tested in accordance with IEC-60068-2-56.) |
Maximum altitude | 2,000 m |
Pollution Degree | 2 |
Indoor use only.
Operational shock 30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC 60068-2-27. Test profile developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating | 5 Hz to 500 Hz, 0.3 grms |
Nonoperating | 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) |
Safety |
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
• EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
• EN 55011 (CISPR 11): Group 1, Class A emissions
• EN 55022 (CISPR 22): Class A emissions
• EN 55024 (CISPR 24): Immunity
• AS/NZS CISPR 11: Group 1, Class A emissions
• AS/NZS CISPR 22: Class A emissions
• FCC 47 CFR Part 15B: Class A emissions
• ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, and additional information, refer to the Online Product Certification section.
This product meets the essential requirements of applicable European Directives, as follows:
• 2014/35/EU; Low-Voltage Directive (safety)
• 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
• 2011/65/EU; Restriction of Hazardous Substances (RoHS)
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column.
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU Customers At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee. 电子信息产品污染控制管理办法(中国RoHS)
中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/ legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in
this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2015—2016 National Instruments. All rights reserved.
371672E-01 Jun16
[1] For the earlier version of the PCI-6230 (192233B-01), refer to the NI 6230 Specifications document, 371672A-01, January 2006 revision.
[2] When outputting 0 V on power down from the analog output channel.
[3] Stresses beyond those listed under Input voltage protection may cause permanent damage to the device.
[4] In other sections of this document, RTSI refers to RTSI <0..7> for the PCI devices or PXI_TRIG <0..7> for PXI devices.
[5] In the figure, |Va – Vd|, |Vb – Vd|, and |Vc – Vd|. 6 In the figure, |Va – Ve|, |Vb – Ve|, and |Vc – Ve|.