Available Configurations

Each light source part number begins with its Legacy Part Number, the predecessor to these light sources. The rest of the part number tells the lamp power and type of lamp included, along with the type of lamp housing and F/# of the condensor lens assembly mounted to each housing. The part number is explained as follows:

Legacy Part Number - Lamp Power and Type - Housing F/#

Legacy Part Number Lamp Power and Type Research Housing F/#
66479-
66481-
66941-
66942-
66944-
200H-200 W Mercury
500H-500 W Mercury
350H-350 W Mercury
R07-0.7
R1-1
R15-1.5 
R22-2.2

The following 200-500 W Mercury Research Lamp Sources are available. Each Lamp Source includes the lamp, socket adapter, housing, cable and power supply as indicated in the table below.

Model Lamp Socket Adapter Housing Lamp Housing Interconnection Cable Power Supply
66479-200H-R22 6283NS 66144 67001 70051 OPS-A500
66479-500H-R22 6285 66162 67001 70051 OPS-A500
66479-350H-R22 6286 66161 67001 70051 OPS-A500
66481-200H-R1 6283NS 66144 67005 70051 OPS-A500
66481-500H-R1 6285 66162 67005 70051 OPS-A500
66481-350H-R1 6286 66161 67005 70051 OPS-A500
66941-200H-R15 6283NS 66144 66901 70051 OPS-A500
66941-500H-R15 6285 66162 66901 70051 OPS-A500
66941-350H-R15 6286 66161 66901 70051 OPS-A500
66942-200H-R1 6283NS 66144 66902 70051 OPS-A500
66942-500H-R1 6285 66162 66902 70051 OPS-A500
66942-350H-R1 6286 66161 66902 70051 OPS-A500
66944-200H-R07 6283NS 66144 66905 70051 OPS-A500
66944-500H-R07 6285 66162 66905 70051 OPS-A500
66944-350H-R07 6286 66161 66905 70051 OPS-A500

Mercury lamps are excellent sources of high intensity DUV-VIS spectral lines; they are widely used as excitation sources. The systems described here run our 200, 350 and 500 W Hg lamps. You can interchange between lamps by simply changing socket adapters in the lamp housing. These sources consist of:
• Oriel Research Arc Lamp Housings with a condenser and a built-in ignitor as indicated per Model number in the table above
• OPS-A500 Power Supply
• Hg DC arc lamp as specified per Model number in the table above
• 70051 Lamp Housing Interconnection cable

Interchangeable Lamps

The 67001 , 67005 , 66901 , 66902 , and 66905 can properly house and operate any 50-500 W DC arc lamp. If a light source from the above table is purchased, other DC Hg arc lamps can be used simply by changing the socket adapter and lamp inside the lamp housing. The appropriate socket adapter for each lamp is shown in the table below:

200 - 500 W Hg Arc Lamps

Mercury Arc Lamp Lamp Wattage Effective Arc Size (mm x mm) Socket Adapter
6283NS 200 0.6 x 2.2 66144
6286 350 0.8 x 2.7 66161
6285 500 1.1 x 4.1 66162

LS-077aFig. 1 Spectral irradiance of 200 W, 350 W, and 500 W Hg Lamps

Lamp Housing

The housing provides a temperature controlled, stable environment for the lamp, and protects personnel from the harmful UV radiation emitted by the lamp. A number of safety features are incorporated into the housing to prevent accidental exposure and lamp overheating. It also houses the ignitor and collimating lens assembly. A socket adapter holds the lamp, vertically, in the center of the lamp housing. To use another suitable lamp with the existing lamp housing, a different socket adapter must be used. External adjustment knobs allow for alignment of the lamp. A reflector mounted behind the lamp collects the lamp\'s back radiation and redirects it to the condenser, maximizing light output. 

Various lamp housings are available for the 200-500 W Hg Research Arc Lamp Sources. Each lamp housing has a unique condensor lens assembly mounted to it. The optical characteristics per each lamp housing below should be considered when selecting a light source.

The new asphere condensers are corrected optics. The design corrects spherical aberration and lateral chromatic aberration with minimized longitudinal chromatic aberrations.

Lamp Housings Comparison

Series Included Lamp Housing Condenser Type Output Beam Output Beam Size Lens Multiplication Factor* Transmittance Range of Lens Material Size Series
66479 67001 F/2.2, fused silica asphere Collimated 1.3 0.05 200 - 2500 nm 1.5 inch
66481 67005 F/1, fused silica asphere Collimated 1.3 0.11 200 - 2500 1.5 inch
66941 66901 F/1.5, single element fused silica Collimated 1.3 0.06 200 - 2500 1.5 inch
66492 66902 F/1, two element fused silica Collimated 1.3 0.11 200 - 2500 1.5 inch
66944 66905 F/0.7, four element fused silica Aspherab® Collimated 2.7 0.20 200 - 2500 3 inch

Power Supply

The OPS-A500 power supply is dedicated to 200-500 W DC arc lamps. It is rugged, reliable and highly regulated. It operates any of the specified lamps under extremely tight power or current control. In addition to superior performance, this supply has a number of features to facilitate source operation. External control of the OPS-A500 can be done via USB or RS 232 interface.

Mounting

Oriel lamp housings have adjustable height instrument feet. You can remove these and hard mount the housing to inch or metric optical tables and baseplates. The optical axis is 4.0 inches (102 mm) above the table. To mount a housing to a bench or rail, order the Carrier Mounting Kit. The kit includes a clamping carrier.

Safety Considerations

These sources emit dangerous levels of UV radiation and ozone. If you do not need output below ~260 nm, choose an ozone free lamp. If this is not an option, vent the ozone to the outside or use our Ozone Eater™ Ozone Converter. To protect yourself from the dangerous UV, wear Ultraviolet Safety Equipment like protective clothing, eyewear and gloves.