特征

Oriel Solar Simulator Set Up Demonstration


Getting any solar simulator up and running is a straightforward process, as demonstrated in this video.


Sol-UV 系列符合哪些标准?

所有 Sol-UV 模拟器的光谱输出均符合 FDA CFR 201.327 部分、ISO 24444:2010 (e) 第一版及国际 SPF 测试方法(CTFASA/COLIPA/JCIA/CTFA:2006 年 5 月,用于光谱匹配。使用可选滤光片,它可以提供符合 JCIA UVA 标准的输出。

适用标准

FDA:21 CFR 201 部分,标签和有效性测试;非处方人用防晒药物产品。§

辐照空间均匀性

工作区域的辐照均匀性对于实现和保持十分重要。热点可能导致输送剂量误差。Sol-UV 模拟器空间均匀性性能标准设计可尽可能减少热点,并在整个工作区域实现低于 5% 的非均匀性。在工作平面上也保持光谱均匀性。

Colipa Irradiance Response Curve

照明器外壳

Sol-UV 外壳为灯提供了安全的外壳。它配备了安全联锁系统,以确保操作员和系统安全。集成风扇和滤光片吹风装置提供强制空气冷却,以使灯具、光学元件和外壳保持最佳温度。使用无臭氧灯可尽可能降低臭氧暴露。Oriel Sol-UV 模拟器采用了一种经久耐用的单片快门,它的设计使用周期是 100 万次。实际使用周期已经超过 1000 万次。快门的最短曝光时间为 200 ms,可通过触点闭合或逻辑电平输入控制,也可通过照明器外壳上便捷的按钮开关进行控制。

时间稳定性

时间稳定性在尽可能减少剂量误差方面也很重要。Oriel 仪器在设计超稳定电源方面经验丰富,输出光随时间而稳定,可确保灯波动不影响所需的剂量的递送。24 小时的典型漂移量小于 2% (最大辐照度-最小辐照度)/(最大辐照度+最小辐照度)

氙弧灯

所有 Sol-UV 型号模拟器都配备有氙弧灯,具有集成式可变衰减器孔径,因此输出可变,在最大可用太阳能常数下,其输出可在 10-100% 间变化。灯和滤光片的组合可产生符合典型 COLIPA/FDA/ISO 标准的输出。系统保持光谱一致性的平均灯寿命为 1000 小时。使用的滤光片无需维护。更换用滤光片(滤光片损坏时使用)单独出售。

电源

超稳压电源为氙灯提供恒定的电力。可以根据电源的累计小时数监测灯的使用情况。在灯具的额定寿命结束时进行更换,以保持光谱匹配度是十分重要的。如果在超过指定寿命(即 1000 小时)的情况下继续使用,将无法保证灯的输出和光谱匹配度。

UVA 滤光片套件

Sol-UV 中的 Sol-UV-A-F 响应曲线

光生物学应用

SPF 测试(体内或体外)、光致毒性测试、光变过敏测试、光致癌性测试和光遗传毒性测试

维护 Oriel Sol-UV 太阳光模拟器

为了便于测量和调整以维持空间均匀性,Oriel 开发了一个手动均匀性测量工具,以便快速测量和确定性能。使用这些工具可以使 Oriel Sol-UV 太阳光模拟器保持在声明的性能规格内。

Arc Lamp Safety


Arc lamps may emit dangerous levels of UV radiation depending on their source type. Ensure that only authorized personnel are in the vicinty of the source, and wearing the necessary safety equipment such as UV protective eyewear, clothing, and gloves. Lighted UV warning signs, posted outside laboratory doors where UV sources are operating, can prevent accidental exposure. Newport also recommends using an electronic shutter to block the beam when the source is not in use but remains turned on. Alternatively, the beam can be safely contained by using beam tubes or fiber optic cables to deliver the light to a sample. It is recommended to vent the ozone that is produced by some lamp models to the outside or use an Ozone Eater.




The Oriel SOL-UV-2 simulator incorporates a patent pending design to be compliant with FDA CFR Part 201.327, ISO 24444:201(e) First Edition, and the International SPF Test Method (CTFASA/COLIPA/JCIA/CTFA): May 2006 for Spectral Match. Validation is conducted by an independent, certified third party laboratory. Non-uniformity of irradiance is < 5% over the entire 2 inch x 2 inch illumination area. Output power can be varied from 10 – 100% of the maximum available solar constants with a uniquely integrated attenuating device. The Oriel Sol-UV simulators use a black non-reflective finish which minimizes stray light. Safety interlocks prevent inadvertent exposure to UV light.